SN204

METHODS OF CONTROL AND CALIBRATION

This reference material cement allows the control of measuring devices in chemistry and particle size of cement according to the test standard CEN EN 196-6. For the other standards, these reference cements are very useful for regular checks to identify a drift.

DELIVERY

The reference cement is offered in 40g bottles.

Each cement is delivered with its analysis certificate

The certificate of analysis was carried out and checked by the offices of France-Ciments in Clichy

PARTICIPATION AND EXECUTION OF THE TESTS

An Inter laboratory testing campaign is organized each year by France-Ciments formerly ATILH. Participating in this campaign are the laboratories of the cement industry in France, Europe and more widely in the world, the laboratories of cement users and research and control centers in the field of construction materials.

This participation is mandatory for laboratories accredited by Cofrac for tests on cement. The tests are carried out according to standard methods where they exist and according to current and traditional methods where they do not.

STATIC ANALYSIS

Outlier removal is performed at the 98% level by the student test. Reiteration is set at this threshold, to keep only the values related to the “Normal or Gaussian” distribution and fully defined by the two parameters: mean and standard deviation. The coefficient of variation symbolized by ‘V’ is the ratio between the standard deviation and the mean ‘X’

SPECIFIC SURFACE

For the calibration of Blaine’s device, follow the requirements of standard NF EN 196-6 for any temperature corrections. To determine the volume of the compacted layer, it is not essential to use the reference cement, but it must be used systematically:

  • After 1000 tries.
  • In cases using another type of pressure fluid, another type of filter paper, a new gauge tube or a new perforated disc.
  • In cases of systematic deviations of the secondary reference cement.

Additional information

Type

Reference Material Cement

Type of Control

Chemical and Granulometry

Type of Material

Composite

Detail of Material

Limestone

Conditionnement

40g Bottle